- High efficiency over 82%
- EN 61000-3-2 Compliance
- Low leakage current
- Low Ripple and Noise
- Short circuit , Overload , Over voltage
Protected
- Input
Voltage : 85VAC~265VAC
- Output Rating
(at 50șC)
+5V and +12V combine power max 120W
,
Total output power max is 130W (ACE-713APM)
+5V and +12V combine power max 185W ,
Total output power max is 200W (ACE-720APM)
+5V and +12V combine power max 265W ,
Total output power max is 280W (ACE-728APM)
- Ripple
and Noise : Less than 1% of each
output at rated load
- Load
Regulation :
- Hold-up
time : 20ms min
- Efficiency:
Higher than 82%
- Protection:
Over voltage/Load protection and short
circuit protection
- Altitude:
0~10,000 ft
- TTL
signal compatible power good signal
- MTBF:
-ACE-713APM: 143,781 hours
-ACE-720APM: 97,572 hours
-ACE-728APM: 97,254 hours
- Vibration:
10-55Hz at 2G , 3min period , 30min along
X,Y and Z axis
- Leakage
current: Less than 500uA at 240V AC
- Shock:
10G for 11ms once on each X,Y and Z axis
- Safety:
UL 2601-1 , IEC 601-1 , EN 60601-1, CCC
- EMI:
FCC docket 20780 curve "B", EN 55011 class "B",
EN 61000-3-2 class "D"
- EMC:
EN 61000-4-2 4KV contactș F8KV air discharge criterion B
EN 61000-4-3 10V/M with 80% AM criterion A
EN 61000-4-4 2KV criterion B
EN 61000-4-5 Line to Line 1KV ; Line to ground 2KV criterion B
EN 61000-4-6 10V with 80% AM criterion A
EN 61000-4-8 30A/m criterion A
EN 61000-4-11 30% dips 10ms , criterion B
60% dips 100ms , criterion C
>95% dips 5000ms , criterion C
-
Operating Temp : 0~70șC
- Storage Temp
: -20~85șC(ACE720APM/ACE728APM)
-40~75șC(ACE730APM)
|
 |
 |

ACE713APM ACE720APM
ACE728APM
|
 |

-
Dimensions :
89(W)x39(H)x152.4(D)mm(ACE713APM)
97(W)x40.5(H)x198(D)mm(ACE720APM)
112(W)x51(H)x203(D)mm(ACE728APM)
Dimensions(ACE713APM):

Dimensions(ACE720APM):

Dimensions(ACE728APM):

| ACE713APM |
130W Medical ATX Power Supply with PFC
function |
| ACE720APM |
200W Medical ATX Power Supply with PFC function |
| ACE728APM |
280W Medical ATX Power Supply with PFC function |
|